Ahwaz Electromechanical Technology

Focusing on Instrumentation: Competitive Pricing with Swift Delivery

Rosemount0085PipeClampTemperatureSensorand248WirelessTransmitter-AhwazAutomation|ProcessControls|AhwazMeasurementInstruments

Temperature Transmitters/Temperature Sensors/Temperature Measurement/ROSEMOUNT/

Rosemount 0085 Pipe Clamp Temperature Sensor and 248 Wireless Transmitter

Specifications Rosemount 0085 Pipe Clamp Temperature Sensor Pipe Clamp Materials: ASTM 304 SST (1.4301), Duplex F51 (1.4462), Carbon steel (1.0037) Lead Wires: PTFE insulated, silver-coated copper wire Insulation Resistance: 1,000 MΩ minimum insulation resistance when measured at 500 VDC at room temperature Calibration Options: Class A Sensor from -50 to 300 °C (-58 to 572 °F) Certifications/Approvals: Hazardous location certifications available; refer to full specifications for complete list. Rosemount 248 Wireless Temperature Transmitter Input: Single sensor capability with universal sensor inputs (RTD, T/C, mV, ohms) Output Signal: WirelessHART protocol Housing: Field mount Display/Interface: Large LCD display Diagnostics: Basic diagnostics including open/short sensor diagnostics Calibration Options: Transmitter-sensor matching (Callendar-Van Dusen constants), 5-point calibration Certifications/Approvals: Hazardous location, marine type; refer to full…

  • Product Details

Specifications

Rosemount 0085 Pipe Clamp Temperature Sensor
Pipe Clamp Materials: ASTM 304 SST (1.4301), Duplex F51 (1.4462), Carbon steel (1.0037)
Lead Wires: PTFE insulated, silver-coated copper wire
Insulation Resistance: 1,000 MΩ minimum insulation resistance when measured at 500 VDC at room temperature
Calibration Options: Class A Sensor from -50 to 300 °C (-58 to 572 °F)
Certifications/Approvals: Hazardous location certifications available; refer to full specifications for complete list.
Rosemount 248 Wireless Temperature Transmitter
Input: Single sensor capability with universal sensor inputs (RTD, T/C, mV, ohms)
Output Signal: WirelessHART protocol
Housing: Field mount
Display/Interface: Large LCD display
Diagnostics: Basic diagnostics including open/short sensor diagnostics
Calibration Options: Transmitter-sensor matching (Callendar-Van Dusen constants), 5-point calibration
Certifications/Approvals: Hazardous location, marine type; refer to full specs for a complete list of certifications
Update Rate: 1 sec. to 60 min., user selectable
Power Module Life: Up to 10-year life, field replaceable (order separately)
Antenna Type: Internal

Features

Rosemount 0085 Pipe Clamp Temperature Sensor
  • Sensor covers wide range of temperatures -328 to 572 °F (-200 ° to 300 °C) to meet application demands
  • Direct mounts to Rosemount 648 Wireless with Rosemount X-well Technology for accurate process temperature measurement
  • Wide variety of clamp styles, sizes, and materials allows for installation anywhere and provides application flexibility
  • Installation requires only two bolts (no welding required) for a straightforward connection
  • Platinum RTD temperature sensors designed with silver or nickel tip for improved response time
  • Withstands applications with high velocities, heavy liquids or particulates for durability and stability
  • Non-intrusive design eliminates need for process penetration of fluids with toxic properties for added safety
  • Spring-loaded sensor ensures best surface contact for enhanced performance
Rosemount 248 Wireless Temperature Transmitter
  • WirelessHART technology is secure, cost-effective and delivers >99% data reliability
  • Designed with an internal antenna for a compact and lightweight design
  • Durable polymer housing offers superior resistance to process environments
  • LCD displays process variables, units, percent of range and diagnostics for increased process visibility
  • SmartPower module provides up to 10-year maintenance-free operation and field replacement without transmitter removal
  • Easy installation enables quick instrumentation of measurement points without the cost of wiring
  • Device dashboards provide easy interface for simplified device configuration and diagnostic troubleshooting
  • Transmitter-sensor matching with Callendar-Van Dusen constants reduces total measurement error

Prev:

Next:

Inquiry

    PersonalBusinessDistributor

    Get a Quote

    Error: Contact form not found.